E2 Pulse Oximeter


This equipment is mainly important for breathing problems as in Breathing problems of a neonate and young infants.

Subsection of this chapter:

  1. Clinical Problem
  2. Assessment
  3. Management
  4. Infection Prevention
  5. Complications
  6. Care & Maintenance
  7. Troubleshooting & Repair


NEST360°. Newborn Essential Solutions and Technologies-Education – Clinical Modules: Pulse Oximeter. (June 2020). License: CC BY-NC-SA 4.0.

Clinical Problem

Assessment of oxygen saturation with a pulse oximeter should be used as part of routine assessment for all infants on admission.

Pulse oximeters should also be used during treatment for all sick or at-risk patients, or those being treated with oxygen therapy (1.1a, 1.1b), CPAP (1.2), or any form of assisted ventilation.(1.3)

Assessment

Peripheral pulse oximetry is a non-invasive & painless process of measuring heart rate and oxygen saturation (oxygen bound to haemoglobin in the capillaries).

Pulse oximeters may be used to help determine the severity of an infant’s illness by evaluating if blood oxygen saturation is low and if respiratory support is needed. Pulse oximeters also may be used to assess the success of treatment and determine a need for increasing or decreasing respiratory interventions to achieve target SpO2.

Pulse oximeters have one red- and one infra-red light-emitting diode and a photodetector. (2.1) The light emitted by the diodes is absorbed by tissues, and amount of absorption is measured by the photodetector. Functional haemoglobin bound with oxygen is called oxyhaemoglobin. Percent saturation of oxygen bound to haemoglobin is called SpO2. A microprocessor within the pulse oximeter determines the percent of oxygen in the blood by comparing the concentration of deoxyhaemoglobin to oxyhaemoglobin at two different light wavelengths. (2.2)

When using a pulse oximeter, several factors impact the stability of the patient’s trace. Examples of a “normal signal” & potential “poor traces” are below. (2.3)

Pulse oximeters may be:

The same pulse oximeter can be used for adult, paediatric, and neonatal patients. The alarm settings should be changed on the oximeter and appropriately sized probes must be used. Probe size will vary depending on patient age.

Normal SpO2 for neonatal patients should be:

If SpO2 readings are less than 90%, the patient should be considered for supplemental oxygen therapy (see oxygen concentrator module).

Oxygen saturations, heart rate, and clinical condition should all correspond.

? ALERT 2.1 Oxygen targets in newborns
Exact oxygen saturation targets for premature newborns remains an
area of controversy. However, most authorities agree that saturations
between 90-95% are reasonable to minimise complications associated
with low and high oxygen levels, including death, neurodevelopmental
impairment, and Retinopathy of Prematurity.1-4

Management

Management of a pulse oximeter covers how to use the device in a variety of settings, including set up for a patient, patient preparation & commencement, care whilst on the device & removal of the patient from the device.

SETTING UP FOR A PATIENT

  1. Follow hand washing procedures.
  2. Collect:
  3. Turn on pulse oximeter by pressing and holding the power button. (3.1) The display should turn on.
  4. Connect the probe. (3.2)
  5. Check for a red light on the probe. If the probe displays a red light, take steps to prepare patient for device. If the probe does not display a red light, follow the guidelines in Pulse Oximeter: Troubleshooting & Repair | If the pulse oximeter is turning on, but no trace is showing.
  6. Clean the pulse oximeter probe thoroughly using alcohol and a cotton swab.
? ALERT 3.1: Low level disinfection solutions
Disinfection of equipment should always comply with manufacturer
guidelines. General guidance on environmental cleaning and disinfection
of equipment was taken from Infection Prevention and Control:
Reference Manual for Health Care Facilities with Limited Resources,
Jhpiego
, 5 Module 6 which lists isopropyl alcohol (70-90%), sodium
hypochlorite (0.05% or >100ppm available chlorine), quaternary
ammonium, and Iodophor germicidal detergent as appropriate for low
level disinfection. Phenolic germicidal detergent is also listed in this
category but should not be used in neonatal wards since affordable,
effective alternatives are available; and, there are concerns it may cause
hyperbilirubinemia and/or neurotoxicity in neonates. 6

PREPARING A PATIENT

  1. Follow hand washing procedures.
  2. Always explain the purpose, risks, and benefits of a procedure to guardians BEFORE performing the procedure.
  3. Collect:
  4. Select best location on patient to collect reading:
  5. Clean location with alcohol & gauze.
  6. Position the patient in a neutral position (3.3) to ensure airway patency.

STARTING A PATIENT

  1. Adhesive (one-use) wrap probe:
  2. Rubber (reusable) wrap probe: (3.4)
  3. Clip probe:

Wrap and clip probes should be firmly placed without need to be held. The probe should not be so tight that it is causing pressure on the skin or impairing circulation.

CARING FOR A PATIENT

  1. Allow the patient’s trace to establish before reading SpO2 and heart rate. (3.8) These should correspond to patient’s clinical condition. If they do not correspond, reposition the probe to ensure good contact with the patient.
  2. Record SpO2 and heart rate in patient documentation. Pulse oximeters are inaccurate for readings under 70%; readings between 20% and 60% do not correlate to clinical deterioration or improvement. (Alert 3.2) A low reading should alert you to look for a problem with the probe fixation, baby or oximeter.
  3. If continuously monitoring patient, periodically check the sensor site (3.9) during monitoring for evidence of skin damage.
  4. Keep the parents informed of baby’s progress.
? ALERT 3.2 Accuracy thresholds for pulse oximeters
WHO Technical Specifications for Oxygen Devices7 lists
accuracy between 70-100% SpO2 as a minimum requirement for all types of pulse
oximeters. Additionally, during NEST360 ̊ technical testing and review,
most (if not all) devices were only tested by the manufacturer to
guarantee accuracy within a certain precision between 70 -100%.

REMOVING A PATIENT

Removing the probe from the patient varies based on the type of probe in use:

  1. If using an adhesive wrap probe: peel adhesive away from patient & pull probe away from patient. Disinfect probe site on patient & wrap probe with 70% alcohol if reusing.
  2. If using a rubber wrap probe: unthread rubber connecting strip through the hole. Pull probe away from patient. Disinfect probe site on patient & the wrap probe with 70% alcohol.
  3. If using a clip probe: press on the tips of the clip probe to open. Gently pull away from patient. Disinfect probe site on patient and clean probe with 70% alcohol.

Infection Prevention

Routine and adequate cleaning of medical devices is critical to prevent hospitalacquired infections in newborn care units. If devices and equipment are not disinfected or reprocessed promptly or adequately between patients, they may pose a significant infection risk.

GENERAL INFECTION PREVENTION

  1. Clean hands with soap and water or 70% alcohol before and after assessing a patient using a pulse oximeter or handling any probes that will be used on a patient.
  2. Ensure that all patient-related consumables (including probes) are new or have been cleaned thoroughly before use. Any patient-related consumables must be cleaned before they are used to assess another patient using the pulse oximeter.
  3. All patient-related consumables should be stored in a clean, dry location. Keep cleaned probes separate from those waiting to be cleaned. Any cables should be loosely wrapped & secured, preventing sharp bends, pinches or kinks, which will decrease their lifetime.

DISINFECTION AFTER USE

  1. Clean reusable probes with alcohol. Adhesive probes are specified for single-use; if reusing disinfect sensor with alcohol. (Alert 4.1)
  2. If pulse oximeters or patient consumables (including probes) are not cleaned thoroughly before use, infection can be transmitted. Care should be taken particularly for consumables marked as single-use but are reused (such as adhesive wrap probes).
  3. Between patients, wipe down the pulse oximeter with alcohol. (4.1) Be careful not to submerge or drip alcohol onto the pulse oximeter or any of its cables.
! ALERT 4.1
While many pulse-oximeter probes are designed to be single use
devices, cost and logistical constraints make this unrealistic in many
low resource settings. Recommendations for cleaning single use
devices were taken from Infection Prevention and Control:
Reference Manual for Health Care Facilities with Limited Resources,
Jhpiego.
5 Vigilance by healthcare workers to assess that pulse-
oximeter readings correlate to patient’s heart rate and clinical
condition is especially critical when re-using probes meant for
single-use. It is not known how rapidly pulse-oximeter probes
degrade or become inaccurate with re-use.

Complications

Introduction of equipment in newborn care units poses clinical and device complications for patients. Awareness of potential complications is critical to maximise patient safety.

CLINICAL COMPLICATIONS


Zoom in to see details

DEVICE COMPLICATIONS


Care & Maintenance

Users are responsible for basic first-line care and maintenance to ensure equipment lasts to their potential lifetime.

POWER SOURCE

A pulse oximeter is powered by replaceable or rechargeable (6.1) batteries. If using a rechargeable device, the users should regularly charge the pulse oximeter when not in use to ensure power in the event of a power outage.

WARD LOCATION

When using the pulse oximeter, the device should be placed in a secure location to prevent drops and breakages. The device and associated probes should be stored in a clean, dry, and secure area. The device should never be placed inside a cot or bassinet.

USER PREVENTIVE MAINTENANCE

Preventive maintenance should be conducted to ensure that the pulse oximeter is in good working order for emergency use.

  1. Turn on the pulse oximeter. Connect the probe and check for a red light.
  2. Connect a clip probe and test readings on your finger for normal saturations (above 90%). (6.2)

Troubleshooting & Repair

Although users are not responsible for repairing their devices, there are steps that may be taken to troubleshoot first-line errors that may occur before contacting maintenance or engineering support.

1 The pulse oximeter is not turning on:
  • Check the batteries or charge on the pulse oximeter.
  • Install new batteries and try again, spare batteries should be available on the ward.
  • Make sure you press and hold the power button for at least 5 seconds.
  • If the pulse oximeter is still not turning on, contact your maintenance department for assistance.

2 Pulse oximeter is turning on, but no trace is showing:
  • Check the probe for a flashing red light. If there is no light, check that the pulse oximeter probe is still connected to the pulse oximeter.
  • If the pulse oximeter probe is still connected and no light is showing, try replacing the probe (and adapter, if using) with a different probe (and adapter, if using).
  • If after replacing the probe (and/or adapter) there is still no light, contact your maintenance department for assistance.

3 If there is a red light flashing but no trace is showing:
  • Wait at least one minute for signal to improve.
  • Ensure that the probe is dry and clean, skin is dry and that the extremity used for assessment is warm and well perfused. Try an alternate extremity.
  • If after addressing above issues you are still unable to obtain a trace contact your maintenance department for assistance.

4 If pulse oximeter is turning on but taking time to stabilise trace:
  • Check that no powerful light sources are shining on the pulse oximeter probe.
  • Confirm that the patient is not moving and that the probe is still securely attached.
  • Confirm the probe is dry and clean.
  • Choose an extremity that is warm, dry, well perfused and wait at least one minute for the signal to stabilise before trying an alternate extremity.
  • If after addressing above issues you are still unable to obtain a trace contact your maintenance department for assistance.

All pulse oximeters have a users’ manual. These should be accessible online. If not available for download online, contact the manufacturer to request access to a copy.


References

  1. Bancalari, E. & Claure, N. Oxygenation Targets and Outcomes in Premature Infants. JAMA 309, 2161 (2013).
  2. Cummings, J. J., Polin, R. A. & Committee on Fetus and Newborn. Oxygen Targeting in Extremely Low Birth Weight Infants. Pediatrics 138, e20161576 (2016).
  3. Manja, V., Lakshminrusimha, S. & Cook, D. J. Oxygen Saturation Target Range for Extremely Preterm Infants: A Systematic Review and Meta-analysis. JAMA Pediatrics 169, 332 (2015).
  4. Polin, R. A. & Bateman, D. Oxygen-Saturation Targets in Preterm Infants. New England Journal of Medicine 368, 2141–2142 (2013).
  5. Curless MS, Ruparelia CS, Thompson E, and Trexler PA, eds. 2018. Infection Prevention and Control: Reference Manual for Health Care Facilities with Limited Resources. Jhpiego: Baltimore, MD.
  6. Sharma, G. Infection Prevention and Control at Neonatal Intensive Care Units. 134.
  7. WHO-UNICEF technical specifications and guidance for oxygen therapy devices. (World Health Organization and the United Nations Children’s Fund (UNICEF), 2019).